Brookfield CAP2000+锥板粘度计提供扩展的温度和粘度范围,具有变速性能。CAP2000+ 还提供较低的剪切率,适用于需要小量样品和良好温度控制的多种场合。坚固耐用和易于使用设计使其完美适应其他仪器所不能适用的环境。
特性和优点
●扩展范围
1.粘度:0.2 - 15,000 泊
2.剪切率:10 - 13,000 sec-1
3.温度:L型 5-75°C 或 H型 50-235°C
4.转速:5-1000 RPM
●满足行业标准:ASTM D4287、ISO 2884 和 BS 3900
●样品量小,< 1 mL,简化测试
●内置式温度控制
●可选CAPCALC32软件,用于完全控制和数据分析
●开放式系统,清洁方便快捷
仪器组成
●仪器主机
●扭矩范围选择:
- 标准扭矩(ICI规格标准):181,000 dyne· cm
- 低扭矩:7,970 dyne· cm
一支用户选定的锥型转子
●温度控制选择(L或H)
MODEL SELECTION: CAP 2000+ Viscometer
|
Viscosity Range
|
SPEEDS
|
MODEL
|
Min.
|
Max.
|
RPM
|
Number of Increments
|
CAP 2000+
|
See chart below
|
5-1000
|
995
|
CAP 2000+ HIGH Torque Cone Viscosity Ranges (Poise) Spindles 01-05
|
Cone Spindle:
|
CAP01
|
CAP02
|
CAP03
|
CAP04
|
CAP05
|
Sample Volume ( μ L) :
|
67
|
38
|
24
|
134
|
67
|
Shear Rate (sec-1):
|
13.3N
|
13.3N
|
13.3N
|
3.3N
|
3.3N
|
CAP 2000+ @ 5-1000 rpm
|
.2-375
|
.4-750
|
.8-1.5k
|
1-3K
|
3-6k
|
CAP 2000+ LOW Torque Cone Viscosity Ranges (Poise) Spindles 01-05
|
Low torque is used for applications requiring low shear rates for low/medium viscosity fluids, an optional low torque 797-7,970 dynes•cm instrument can also be ordered.
|
CAP 2000+ @ 5-1000 rpm
|
.2-16
|
.2-32
|
.2-66
|
.2-130
|
.2-260
|
CAP 2000+ HIGH Torque Cone Viscosity Ranges (Poise) Spindles 06-10
|
Cone Spindle:
|
CAP06
|
CAP07
|
CAP08
|
CAP09
|
CAP10
|
Sample Volume ( μ L) :
|
30
|
1700
|
400
|
100
|
170
|
Shear Rate (sec-1):
|
3.3N
|
2.0N
|
2.0N
|
2.0N
|
5.0N
|
CAP 2000+ @ 5-1000 rpm
|
8-15k
|
.78-625*
|
3.13-2.5k*
|
12.5-10k*
|
1-1k*
|
CAP 2000+ LOW Torque Cone Viscosity Ranges (Poise) Spindles 06-10
|
Low torque is used for applications requiring low shear rates for low/medium viscosity fluids, an optional low torque 797-7,970 dynes•cm instrument can also be ordered.
|
CAP 2000+ @ 5-1000 rpm
|
.2-660
|
.2-26
|
.2-108
|
.2-440
|
.2-44
|
μL = microLiter N = rpm k = 1000 e.g. Cone CAP-01 13.3 x 10 (rpm) = 133 sec-1
* Maximum speed recommended with this spindle is 400 rpm. Viscosity range indicated is for operation at 400 rpm.
Note: Viscosity ranges shown are for illustration; the exact range will depend upon instrument configuration.